Used for measuring magnetic emissions to specification limits in MIL-STD-461, 462, 463 and now in other specifications and standards. It is generally used connected to an EMI receiver covering the required frequency range. Typical compatible receivers would be items 105;120;121;116. Not for transmitting.
Electrical Specifications:
Useful Freq Range: 20Hz to >500kHz.
Conversion Factor: Refer to handbook.
Coil Diameter: 13.3cm.
Winding: 36 turns of 7-41 Litz.
Resistance: 10Ω, nominal.
Inductance: 340 uH, nominal.
Connector: BNC female.
Mechanical:
Case Dimensions 27cm x 22cm x 7.5cm. (LxWxH):
Case weight: 0.6Kg