Search:
Application Area:EM Transducers Radiated (300's)Product Area:Low Frequency Magnetic FieldFrequency Range:20Hz - 50kHzTuning:NoneInterface:BNC (f) & 4mm skts.Other:N/AEMC Item No:310
This magnetic field generating coil is used in Magnetic Susceptibility Tests for qualifying equipment to specification MIL-STD-461D & 462D. method RS01. It is built in rigorous conformance with the description in that specification and is made of Delrin. When RS01 is being performed, a 1Ω, 100W resistor should be connected in series to monitor current by a suitable meter. (item 267B). However, some aspects of Mil.Std. 462D require a monitoring loop to actually confirm the strength of the magnetic field at a specific point. This monitoring loop can be found under (item 317). See inset photo.
Specifications:
Electrical:
Coil: 10 turns of No 16 SWG wound on a non- conducting Delrin form. Useful Frequency range: 20Hz - 50kHz ResonantFrequency:>100kHz
Maximum Current: 10 amps short time, 5 amps continuously
Coil Constant: 5 x 10-5 Tesla/amp at 5cm from wire turns
1 Tesla = 1 Weber/sq. meter = 104 gauss
Mechanical:
2 female banana jacks spaced: 0.75” apart
Nominal Coil Diameter: 12cm (4.75”)
Thickness of Coil: 0.3cm
Weight: 0.45kg