Application Area:EM Transducers Radiated (300's)Product Area:Low Frequency Magnetic FieldFrequency Range:DC - 2kHzTuning:NoneInterface:N/AOther:Calibration Graph suppliedEMC Item No:313
Model 6403 Helmholtz coil system has twin 0.9 metre (3ft) diameter coils in a Helmholtz configuration and normally used as a Magnetic Field Immunity Test System or simply a low frequency magnetic field calibrator. A magnetic field is produced by two coils that is essentially homogenous throughout the volume between the face of the coils and one-third diameter of the coils at it’s centre. The coils are supported by a non-metallic framework. A rotating, height-adjustable pedestal is capable of holding articles weighing up to 34kg. The Coil has been initially calibrated at the NPL in terms of uT/amp at 50Hz. For the end user’s convenience NPL have also quoted a figure of uTesla per millivolt as measured across an installed shunt resistor with measuring terminal placed at the base of the Helmholtz. This was done at 60Hz and 2kHz. Typical examples of field using the crown LF amplifier can be seen on the graph where the Crown amplifier (item 402) is hard driven with 3.1vrms input. At 1kHz, a field of 57.3 u/Tesla is achievable with the amplifier supplying 56mA (hard driven) and at 60Hz & a drive of 8.8 ampers, a field of 1m/T was achieved. Built in on the base is the feed connector with 1Ω resistor for current verification. Also a 4metre amplifier o/p supply lead. The device can arrive disassembled or assembled.
Specifications:
Electrical:
Coil Factor: 118.7 uTesla/ampere
Homogeneity: ± 5% within a 30cm cube
Coil Impedance: Approx. 0.5ohms DC;
Inductance: 18.4 millihenrys
Self Resonance: 18kHz
Max I Input: 15 amps continuous
** 1 gauss = 100µT = 160dBµT = 80 Amps/metre
Mechanical:
Diameter: 0.9m (3’)
Weight: approximately 24kg (53lbs).