EMC Item no.: | 955 |
Category: |
|
Test Mode: | Immunity |
Test Type: | System |
Frequency: | 50kHz-1GHz |
During Conducted Immunity (CI) testing, RF voltages or currents are applied to the cables connected to the Equipment Under Test (EUT) in order to confirm that the RF does not degrade performance or cause deviations from the specifications of the EUT.
An integrated Conducted Immunity system (and its calibration jig) is included along with a signal generator, amplifier, coupler, power meter and probe, as well as RF cables with terminations suited to the devices provided. This bespoke system can be used for non-standard levels when testing against EN 61000-4-6, or any other RF conducted immunity test.
EMC Hire engineers will select equipment based on the customer's needs, so please provide as much information as possible (e.g. applicable standard, test level, size of EUT cables, etc.).
The injection device can be CDN, EM Clamp, BCI probe, Coupling Transformer depending on the standard.
A monitoring probe can be used, depending on the standard applied, to record the amount of current flowing in the cable under test.
A laptop containing RadiMation test software is included in the system, which will be preloaded with the appropriate test configurations including transducer factors. A software operating procedure is also included to guide you through using the software.
We invite you to visit our facility for a demonstration and training, or EMC Hire can provide an engineer to set up the system at your location if necessary, or provide remote assistance via Teamviewer or Teams.
The specification of the test equipment for this group is tailored around the customer's requirement.
Frequency ranges and test levels vary depending on the standard applied, our engineers will advise on the best equipment to meet your requirements.
The table below shows some of the available options for the Conducted Immunity system. Include the option number in your enquiry to aid in the item selection.
Option No. |
Test Method |
Applications |
#1 |
EN 61000-4-6 |
Signal Ports |
#2 |
DC Power Ports |
|
#3 |
AC Power Ports |
|
#4 |
ISO 11452-4 |
BCI Test Method - Substitution |
#5 |
BCI Test Method - Closed Loop |
|
#6 |
Def Stan 59-411: DCS02 |
Air, Land & Sea Service |
#7 |
Mil Std 461: CS114 |
Air, Land & Sea Service |
In response to your enquiry via phone or through this website, an engineer will contact you to discuss your requirements. We advise that you download the below Test Systems Enquiry Form to expedite the process and ensure that your requirements are fully understood and met.